Events News Sitemap INTEGRA Holding Deutsch
 
Technical data
 
 
 
 
 
MediaClave - Technical Data
Max. content of pressure chamber
12 Litre  
Diameter and usable chamber height when autoclaving
Ø 269 mm, H 190 mm
Amount of agar that can be prepared min./max.
1 - 9 Litre
Temperature range
70°C - 122°C
Temperature range when using as water bath
37°C - 80°C
Sterilisation time
1 - 99 min
Temperature range for removal
30°C - 80°C
Max. temperature deviation
+/- 0.45°C
Heat-up period (25°C -> 121°C) 1- 9 L
12 - 33 min
Cool-down period (121°C -> 55°C) 1 - 9 L
6 - 20 min
Preparation time for 1 - 9 Litre agar
38 - 73 min
Stirrer speed
Stage 1: 75 U/min
Display
Stage 2: 150 U/min
LC- graphic facility 240x64 points
Heating capacity
3.0 kW
Weight
49 kg (60 kg packed)
Dimensions h x l x d
437 x 570 x 490 mm
(760 x 600 x 600 mm packed)
Water connections
3/4"
Storable programs
Sterilisation of media, freely selectable: 13
- Preparation of chocolate agar: 1
- Preparation of sensitive media (105ºC),
  predefined : 13
  Autoclave mode, freely selectable: 15
  Water bath mode, freely selectable: 15

 

 
  © INTEGRA Biosciences AG - Design by aaa-design.ch >  back to top