Specifications VIAFLO Pipette

VIAFLO Electronic Pipettes Specifications

VIAFLO, Single Channel

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

4011

0.5 – 12.5

0.01

1.25
6.25
12.5

5.00
1.50
1.00

4.00
0.80
0.40

4016

2 – 50

0.05

5
25
50

3.00
1.50
1.00

1.50
0.60
0.40

4012

5 – 125

0.1

12.5
62.5
125

3.00
1.20
0.60

1.00
0.40
0.20

4013

10 – 300

0.5

30
150
300

2.00
1.00
0.60

0.60
0.30
0.15

4014

50 – 1250

1

125
625
1250

3.00
1.00
0.60

0.60
0.30
0.17

4015

100 – 5000

5

500
2500
5000

3.00
1.20
0.60

0.75
0.30
0.15

VIAFLO, 8 Channel

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

4621

0.5 – 12.5

0.01

1.25
6.25
12.5

10.00
4.00
2.00

6.00
1.60
0.80

4626

2 – 50

0.05

5
25
50

5.00
2.50
1.50

2.50
0.60
0.40

4622

5 – 125

0.1

12.5
62.5
125

3.75
2.50
1.60

1.50
0.70
0.35

4623

10 – 300

0.5

30
150
300

4.00
2.00
1.60

1.20
0.60
0.35

4624

50 – 1250

1

125
625
1250

6.00
2.40
1.20

1.10
0.50
0.30

VIAFLO, 12 Channel

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

4631

0.5 – 12.5

0.01

1.25
6.25
12.5

10.00
4.00
2.00

6.00
1.60
0.80

4636

2 – 50

0.05

5
25
50

5.00
2.50
1.50

2.50
0.60
0.40

4632

5 – 125

0.1

12.5
62.5
125

3.75
2.50
1.60

1.50
0.70
0.35

4633

10 – 300

0.5

30
150
300

4.00
2.00
1.60

1.20
0.60
0.35

4634

50 – 1250

1

125
625
1250

6.00
2.40
1.20

1.10
0.50
0.30

VIAFLO, 16 Channel

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

4641

0.5 – 12.5

0.01

1.25
6.25
12.5

10.00
4.00
2.00

6.00
1.60
0.80

4646

2 – 50

0.05

5
25
50

5.00
2.50
1.50

2.50
0.60
0.40

4642

5 – 125

0.1

12.5
62.5
125

3.75
2.50
1.60

1.50
0.70
0.35

Remarks

  • The multichannel specifications represent the performance of all channels of a pipette.
  • Precision = Coefficient of variation

 

Download 

Specifications VIAFLO Pipette as PDF