3 simple and proven automation protocols for serial dilution on the ASSIST PLUS pipetting robot

  • 3 simple and proven automation protocols for serial dilution on the ASSIST PLUS pipetting robot

    Serial dilution made easy

    Serial dilution, a reduction in concentration by constant dilution factor, is a common approach for screening related applications, such as the determination of minimum inhibitory concentrations (MIC) in drug discovery, the most probable numbers (MPN) in microbiology or general nucleic acid quantifications in molecular biology. Despite being a simple technique, poor liquid handling of interdependent dilution steps can cause error propagation and accumulation. Thorough mixing, therefore, is crucial but puts a lot of strain on the thumb, which increases the risk of repetitive strain injuries. Furthermore, when done regularly, serial dilutions are significantly time-consuming procedures.

    Here we show the simplest way of how to do serial dilutions with the VOYAGER adjustable tip spacing pipette on the ASSIST PLUS pipetting robot to gain more walk-away time. The provided protocols detail optimal settings with reliable results when diluting analytes into water. In our comprehensive guide you can learn even more about key parameters when facing different conditions.

  • Table of contents

    Serial dilution, a reduction in concentration by constant dilution factor, is a common approach for screening related applications, such as the determination of minimum inhibitory concentrations (MIC) in drug discovery, the most probable numbers (MPN) in microbiology or general nucleic acid quantifications in molecular biology. Despite being a simple technique, poor liquid handling of interdependent dilution steps can cause error propagation and accumulation. Thorough mixing, therefore, is crucial but puts a lot of strain on the thumb, which increases the risk of repetitive strain injuries. Furthermore, when done regularly, serial dilutions are significantly time-consuming procedures.

    Here we show the simplest way of how to do serial dilutions with the VOYAGER adjustable tip spacing pipette on the ASSIST PLUS pipetting robot to gain more walk-away time. The provided protocols detail optimal settings with reliable results when diluting analytes into water. In our comprehensive guide you can learn even more about key parameters when facing different conditions.

Key benefits

  • Proven serial dilution protocols with optimal settings for the VOYAGER adjustable tip spacing pipette on the ASSIST PLUS pipetting robot guarantee uniform pipetting and mixing.
  • INTEGRA's electronic pipettes prevent thumb overtaxing during liquid handling steps and, together with the ASSIST PLUS, enable risk-free handling of hazardous samples.
  • Profit from additional hands-free time with the ASSIST PLUS and stop wasting time doing simple standard procedures manually.
  • Game-changing liquid handling for 2-, 5- and 10-fold serial dilutions, with dynamic mixing volumes ensuring homogeneity of analytes.
  • Simplified workflow with VIALAB's serial dilution protocol, including individual mixing parameters to cope with poorly soluble analytes.
  • More flexibility during automated serial dilution with the VOYAGER adjustable tip spacing pipette when performing serial dilutions between different tubes and plates. Furthermore, the pipette can be easily changed to another volume size while using the same protocol.

Overview: How to do serial dilution with the ASSIST PLUS pipetting robot

Here we show how to do serial dilution of tartrazine in water with an 8 channel 125 µl VOYAGER adjustable tip spacing pipette on the ASSIST PLUS pipetting robot.

Experimental set-up

The ASSIST PLUS pipetting robot, together with the 8 channel 125 µl VOYAGER adjustable tip spacing pipette and 125 µl sterile, filter GRIPTIPS®, automates complete serial dilutions in one program consisting of three steps (Figure 1):

  1. Transfer diluent to target plate
  2. Transfer analyte to target plate
  3. Serial dilution of analyte within target plate
Illustration of a step-by-step procedure of how to do serial dilutions.
Figure 1: How do serial dilution (experimental set-up).

Downloads: App note and protocols for serial dilution with the ASSIST PLUS pipetting robot

Step-by-step procedure

Graphical representation of the ASSIST PLUS pipetting robot deck set-up with INTEGRA dual adapter for reagent reservoirs, a 96 well flat-bottom plate and the VOYAGER adjustable tip spacing pipette.
Figure 2: Deck set-up for serial dilution. Position A: Source – dual reservoir adapter with two 10 ml reservoir; diluent in A1 (blue) and analyte in A2 (green). Position B: Target – 96 well flat-bottom plate (pink). Position C: Empty.

Serial dilution of an analyte

The INTEGRA dual reservoir adapter together with two 10 ml reservoirs is placed on deck position A, with diluent (blue) in A1 and analyte (green) in A2 (Figure 2). A clear 96 well flat‑bottom plate (pink) is placed in landscape orientation on deck position B (Figure 2).

Select and run one of the following VIALAB programs:

- 2-fold serial dilution → 125_VOYAGER_2_fold_serial_dilution

- 5-fold serial dilution → 125_VOYAGER_5_fold_serial_dilution

- 10-fold serial dilution → 125_VOYAGER_10_fold_serial_dilution

Specific volumes are handled by the VOYAGER adjustable tip spacing pipette (Figure 3). The diluent is transferred in multiple dispensing steps from the reservoir (Position A-A1) into each well of the 96 well flat-bottom plate starting with column two (Position B). For each dispense, a pre- and post-dispense with 5 % (10-fold serial dilutions) or 10 % (2- and 5-fold serial dilutions) of the transferred volume is used to guarantee precision during plate set-up.

Illustration of 2-fold serial dilution, 5-fold serial dilution and 10 fold serial dilutions in 96 well flat bottom plates when using the 125 µl VOYAGER adjustable tip spacing pipette.
Figure 3: How to do serial dilutions in 96 well flat-bottom plates with the 125 µl VOYAGER.

Using new GRIPTIPS, the VOYAGER aspirates the highest concentration of analyte from the reservoir (Position A-A2) and dispenses it into the first column of the 96 well flat‑bottom plate (Position B).

Without changing the GRIPTIPS, the VOYAGER begins the serial dilution by aspirating the specific volume (Figure 3) from column one of the 96 well flat‑bottom plate (Position B) and dispensing into the second column. The VOYAGER mixes then 100 µl (2- and 10-fold serial dilutions) or 80 µl (5-fold serial dilutions) of the analyte/diluent five times at maximum speed (10). Before aspirating for the following dilution step, a blow out is performed to clear the tip of any remaining liquid. The procedure is repeated until column 11 is reached, where the last aspiration is discarded along with the GRIPTIPS (Figure 4). Column 12 only contains diluent, and functions as blank for background noise elimination.

ASSIST PLUS pipetting robot together with the 8 channel 125 µl VOYAGER adjustable tip spacing pipette is performing a serial dilution.
Figure 4: The ASSIST PLUS and VOYAGER performing serial dilution of tartrazine.

Tips:

  • Excellent accuracy and precision with tip pre-wetting when pipetting aqueous liquids.
  • Adjustable mixing cycles to compensate slower mixing speed or poorly soluble analytes.

Results

The performance of the 8 channel 125 µl VOYAGER adjustable tip spacing pipette on the ASSIST PLUS pipetting robot during serial dilution of 0.36 mM tartrazine in water in 96 well flat‑bottom plates (Figure 5) was analyzed at 428 nm absorbance (Tecan Infinite® M200PRO). Detailed data is provided in our comprehensive guide on how to perform serial dilution.

 

A 96 well flat-bottom plate with a finished 2-fold serial dilution of tartrazine.
Figure 5: 2-fold serial dilution of tartrazine in a 96 well flat-bottom plate.

Figure 6 shows a representational, optimized calibration curve of a 2-fold serial dilution. Automating all liquid handling steps and mixing each dilution with 100 µl (>80 % GRIPTIPS volume) five times at maximum speed (10), led to reliable results in three independent runs. Furthermore, final values of below 1 % were calculated for the accuracy and precision of the individual dilution steps.

Illustration of a tartrazine 2-fold serial dilution performed with optimal settings on the ASSIST PLUS pipetting robot. A linear regression plot shows the calibration curve of three independent runs.
Figure 6: Result of 2-fold serial dilution of tartrazine using optimized mixing settings on the ASSIST PLUS.

Remarks

  • VIALAB software: VIALAB programs can be easily adapted to your specific pipette, labware and protocols, for instance when partial plates are needed.
  • Partial plates: Programs can be adapted at any time to a different number of samples, giving laboratories total flexibility to meet current and future demands.

Conclusion

  • Profit from reproducible results and eliminate any operator influence on serial dilutions with automated workflows on the ASSIST PLUS pipetting robot.
  • Risk-free dilution of hazardous compounds in a biosafety cabinet is possible thanks to the compact footprint of the ASSIST PLUS.
  • Guaranteed homogeneity of aqueous solutions with dynamic mixing of each dilution is only possible when using INTEGRA's electronic pipettes (80 % of total GRIPTIPS or reaction volume, five times, speed 10).
  • Understanding how to perform serial dilutions is crucial to optimizing workflows and preventing error propagation. The automated protocols on the ASSIST PLUS provide the best liquid handling settings for 2-fold serial dilutions, 5-fold serial dilutions and 10-fold serial dilutions.

Ask our expert. Leave a comment!

Write us if you have any questions regarding the application note or one of our instruments.

Any questions? I'm happy to help!

  • Rebecca Bodmer

    Application Specialist

    Email

Instruments and accessories

ASSIST PLUS, Pipetting Robot VOYAGER, 8 Channel, 125 µl Dual Reservoir Adapter Divided Reagent Reservoir, Sterile, SureFlo anti-sealing array, 25 ml, PS GRIPTIPS®, 125 µl, Sterile, Filter (for automation systems) Greiner Bio-One: 96 Well Microplate, PS, F-Bottom

ASSIST PLUS, Pipetting Robot

INTEGRA has developed the ASSIST PLUS pipetting robot to streamline routine pipetting tasks at an affordable price. Using INTEGRA electronic multichannel pipettes, the system:

  • automates pipetting tasks,
  • eliminates physical strain and
  • ensures superior reproducibility and
  • error free pipetting.

Part No. 4505

Visit product page
ASSIST PLUS Pipetting Robot

VOYAGER, 8 Channel, 125 µl

VOYAGER pipettes allow the tip spacing to expand anywhere between 4.5 mm and 33 mm at the push of a button.

  • Single handed operation leaves the other hand free to handle labware.
  • On the fly access to up to 3 user-defined tip spacings.
  • These user defined tip spacings are saved. No need to memorize spacings of different labware formats.

Part No. 4722

See it work
VOYAGER pipettes allow the tip spacing to expand anywhere between 4.5 mm and 33 mm at the push of a button.

Dual Reservoir Adapter

The dual reservoir adapter makes this system ideal for assays that required sequential addition of several low volume reagent mixes. This adapter holds two 10 ml, 25 ml or divided reagent reservoirs side by side in an aluminum frame, and can be pre-chilled to keep reagents cool during processing.

Part No. 4547

Find out more about ASSIST PLUS accessories
Dual reservoir adapter with 2 didived reservoirs

Divided Reagent Reservoir, Sterile, SureFlo anti-sealing array, 25 ml, PS

Lowest Dead Volume On The Market!

25 ml divided reservoirs are comprised of 5 and 10 ml compartments, while still nesting comfortably in the standard 25 ml reusable, sturdy base. 
Each reservoir features the SureFlo™ anti-sealing array and hydrophilic surface treatment providing especially low dead volumes.

Part No. 4350, 4351, 4352

Find out more about Divided Reagent Reservoirs
divided-reagent-reservoir-to-well-plate.jpg

GRIPTIPS®, 125 µl, Sterile, Filter (for automation systems)

GRIPTIPS® pipette tips perfectly match the multi-lobe tip fitting, snapping firmly on during loading. This guarantees a perfect seal on every tip, preventing them from loosening, leaking or completely falling off. All tips are precisely aligned horizontally, enabling accurate touch-offs, even when pipetting with 384 tips.

Part No. 6465

Learn more about GRIPTIPS® pipette tips for Benchtop Pipetting Systems
Integra-Relaunch_Website-Output-Produktebilder-Product_Overview-GripTips.png

Greiner Bio-One: 96 Well Microplate, PS, F-Bottom

SAPPHIRE MICROPLATE, 96 WELL, PP, FOR RT PCR, WHITE, HALF-SKIRT, FOR LIGHT CYCLER, 10 PCS./BOX

Source: Website Greiner Bio One

Visit supplier website

Customer’s voice

The ASSIST PLUS is a friendly, easy to use, liquid handling system that helps us perform the routine lab work with ease and avoiding human errors.

Sandeep Kumar, Viral Solutions