-
Automated size selection with CleanNGS – the cost-effective alternative to AMPure XP
Proven NGS size selection protocols on the ASSIST PLUS pipetting robot
DNA size selection with magnetic beads plays a significant role in molecular biology, employing specific ratios to capture and separate fragments by size. After the polymerase chain reaction (PCR), single-sided DNA size selection uses 1 high magnetic bead ratio to remove primer dimers, ensuring PCR product purification for efficient downstream sequencing. In contrast, DNA double-sided size selection uses 2 ratios to effectively remove small and large fragments, resulting in the purification of specific target average fragment sizes. Both single- and double-sided DNA size selections with magnetic beads are integral to next generation sequencing (NGS) library preparation.
Level up your PCR clean-up protocol and DNA size selection workflow using magnetic beads with our fully automated size selection protocols. The VOYAGER adjustable tip spacing pipette on the ASSIST PLUS pipetting robot automates all liquid handling steps and the MAG module ensures precise automated magnetic bead handling. The protocols provided demonstrate accurate handling of magnetic beads for reproducible PCR product purification and DNA size selection. Using CleanNGS and AMPure XP magnetic beads (Beckman Coulter Life Sciences) confirmed their interchangeability, establishing CleanNGS as a cost-effective alternative offering reliable results.
-
Table of contents
DNA size selection with magnetic beads plays a significant role in molecular biology, employing specific ratios to capture and separate fragments by size. After the polymerase chain reaction (PCR), single-sided DNA size selection uses 1 high magnetic bead ratio to remove primer dimers, ensuring PCR product purification for efficient downstream sequencing. In contrast, DNA double-sided size selection uses 2 ratios to effectively remove small and large fragments, resulting in the purification of specific target average fragment sizes. Both single- and double-sided DNA size selections with magnetic beads are integral to next generation sequencing (NGS) library preparation.
Level up your PCR clean-up protocol and DNA size selection workflow using magnetic beads with our fully automated size selection protocols. The VOYAGER adjustable tip spacing pipette on the ASSIST PLUS pipetting robot automates all liquid handling steps and the MAG module ensures precise automated magnetic bead handling. The protocols provided demonstrate accurate handling of magnetic beads for reproducible PCR product purification and DNA size selection. Using CleanNGS and AMPure XP magnetic beads (Beckman Coulter Life Sciences) confirmed their interchangeability, establishing CleanNGS as a cost-effective alternative offering reliable results.
Key benefits
- Efficient automated magnetic bead handling when using the MAG module – our module for magnetic separation. MAG captures and releases magnetic beads with vertical back and forth magnet movement to the plate without user interference, avoiding possible spillage during manual plate transfer.
- Efficient and reproducible PCR product purification for downstream sequencing uses CleanNGS magnetic beads and removes fragments below 100 bp.
- Fail-proof liquid handling of magnetic beads with the VOYAGER adjustable tip spacing pipette on the ASSIST PLUS pipetting robot, thanks to optimized pipetting height and speed settings in VIALAB.
- Reduce the processing costs and boost your NGS library preparation reproducibility for double-sided DNA size selection with CleanNGS magnetic beads.
- Flexibility is key! No matter what labware you are using, MAG's adapter selection provides optimized solutions for automated magnetic bead handling in microcentrifuge tubes, 96 and 384 well PCR plates.
- Gain additional hands-free time with ASSIST PLUS and MAG and effortlessly adapt to various DNA size selection protocols. VIALAB's user-friendly programming makes adjusting magnetic bead ratios easy.
Overview: How to automate DNA size selection with ASSIST PLUS and MAG
In this application note, we demonstrate fully automated DNA size selection of 48 samples with CleanNGS magnetic beads on the ASSIST PLUS pipetting robot. The 8 channel 125 µl VOYAGER adjustable tip spacing pipette automates the liquid handling steps and the MAG module automates the magnetic bead handling steps in this protocol (we provide a semi-automated workflow using a magnet plate in the appendix).
Figure 1 illustrates the step-by-step procedure of the provided size selection protocols for:
- Single-sided DNA size selection (PCR product purification)
- Double-sided DNA size selection
Downloads: App note and protocols for single- and double-sided DNA size selection with the ASSIST PLUS pipetting robot
DNA single-sided size selection (PCR product purification)
Bind PCR product to a 1.8x magnetic bead ratio
Prepare fresh 80 % ethanol, and bring CleanNGS magnetic beads up to room temperature (RT). Place a 96 well DWP in portrait orientation on position A with 450 µl of CleanNGS magnetic beads in wells A1-A8 (Figure 2, light blue), 1.6 ml of 80 % ethanol in wells B1-B8 (Figure 2, blue) and 320 µl of molecular grade water in wells C1-C8 (Figure 2, pink). Place one empty 96 well HardShell PCR plate on position B and one with 40 µl of sample in each well of the first half in landscape orientation on the MAG module with the 96 well adapter on position C (Figure 2, yellow).
Select and run the VIALAB program 'MAG_PCR_product_purification'. With 125 µl sterile, filter, low retention GRIPTIPS®, VOYAGER on ASSIST PLUS will transfer 72 µl of magnetic beads from column A (Figure 2, light blue) of the DWP on position A (Figure 3a) to every well of the first half of the PCR plate on position C (Figure 3b). Magnetic beads are mixed 10 times before aspiration, and 15 times after dispensing into samples, guaranteeing a homogenous magnetic bead mixture. GRIPTIPS are changed automatically between samples. With the magnet array disengaged (Figure 4a, position home, 0 mm), VOYAGER will initiate an incubation of 5 minutes at RT to bind the PCR fragments to the magnetic beads (Figure 4b).
Removal of small fragments and washing
MAG will automatically lift the magnet array from position home (Figure 4b) to position high at 29 mm (Figure 5a) and capture magnetic beads within 3 minutes (Figure 5b). The PCR plate adapter for MAG has small holes to visually confirm targeted capturing of magnetic beads at the well surface.
VOYAGER – using fresh GRIPTIPS for each sample – removes the supernatant while the magnet array remains engaged (Figure 5a). The pipette transfers the supernatant into columns F-H of the DWP on position A. Slow aspiration (speed 1) and precise height settings prevent magnetic bead loss during washing. Magnetic beads are washed twice with 125 µl of 80 % ethanol from columns B of the DWP on position A (Figure 2, blue). VOYAGER aspirates an additional time to ensure complete removal of ethanol from each well. MAG will lower the magnet array by 5 mm (position low, 24 mm) followed by air drying for 3 minutes at RT. Lowering the magnet array before air drying will move the pellets closer to the well bottom, allowing easier elution and smaller volumes.
Tips:
- The magnet step in VIALAB provides total control of the magnet array, by setting customized heights anywhere between 0 and 29 mm.
- Worry-free liquid handling of ethanol uses fast aspiration and slow dispensing with a tip touch to prevent droplet formation.
Elution of single-sided size selected fragments
40 µl of molecular grade water is then transferred from column C of the DWP on position A (Figure 2, pink) to every well of the first half of the PCR plate on position C. Mixing 25 times ensures proper resuspension of the magnetic beads, independently of the volume, followed by a 5-minute RT incubation for elution. Again, MAG will automatically lift the magnet array to 29 mm (Figure 5a) and capture magnetic beads within 3 minutes (Figure 5b). Afterwards, VOYAGER transfers 35 µl of eluate to the unused PCR plate on position B, leaving 5 µl in the plate at position C to prevent magnetic bead carryover. At the end of the run, the user is prompted to store the PCR plate from position B, and remove the plate from the MAG.
DNA double-sided size selection
Binding sheared genomic DNA (gDNA) to a 0.7x magnetic bead ratio
The deck set-up for DNA double-sided size selection is similar to DNA single-sided size selection, but with 320 µl of CleanNGS magnetic beads (Figure 2, light blue), 350 µl of molecular grade water (Figure 2, pink), and 55 µl of sample in each well of the first half of a HardShell 96 well PCR plate (Figure 2, yellow).
Select and run the VIALAB program 'MAG_DNA_double_size_selection'. VOYAGER will follow the steps described in PCR product clean-up but transfers 38.5 µl of magnetic beads to each well containing a sample (Figure 3b). Mixing 10 times before every other aspiration using new GRIPTIPS before aspiration guarantees precise, low volume pipetting of magnetic beads.
Removal of large fragments (right size selection)
After capturing large fragments bound to magnetic beads (right size selection) (Figure 5b), VOYAGER will transfer 85 µl of supernatant from each well of the first half of the PCR plate to the corresponding well of the second half of the same plate on position C.
Binding target fragments to a 0.8x magnetic bead ratio, and removal of small fragments (0.8x, left size selection) during the washing process
Following the same procedure as the right size selection, MAG lowers the magnet array back to position home (Figure 4a) and VOYAGER transfers 5 µl of magnetic beads to the supernatant of the first size selection in the second half of the PCR plate (position C). A 5 µl pre-dispense guarantees accurate pipetting of small volumes of magnetic bead. The subsequent washing procedure mirrors the DNA single-sided size selection.
Elution of double-sided size selected fragments
MAG and VOYAGER will follow the same procedure used for DNA single-sided size selection but transfer 50 µl of molecular grade water before elution, and 45 µl after capturing magnetic beads with MAG (Figure 5b).
Tip:
- Changing to different fragment sizes is trouble free, as the operator can calculate the magnetic bead volume for any ratio, and simply update it in VIALAB.
Results
Most reagent kit providers for library preparation recommend AMPure XP magnetic beads for NGS. Here, we demonstrate equivalent performance of CleanNGS magnetic beads during full automation of single-sided DNA size selection using a 100 bp DNA ladder (Promega) to mimic PCR product purification and double-sided DNA size selection using sheared gDNA.
With the VOYAGER adjustable tip spacing pipette on the ASSIST PLUS pipetting robot, 48 replicates were processed using CleanNGS magnetic beads in rows A to D, and AMPure XP magnetic beads in rows E to H. Automated magnetic bead handling with optimized magnet array heights was ensured with the MAG module. The size-selected fragments (CleanNGS vs AMPure XP) were analyzed and compared using the 4150 TapeStation System (Agilent, complete data set can be found in the appendix).
Figure 6 illustrates the gel picture of row A (CleanNGS) and row E (AMPure XP) of the 96 well plate when performing single-sided DNA size selection with 100 bp DNA ladder and a 1.8x magnetic bead ratio. Both magnetic beads purified all fragments of PCR ladder smaller than 100 bp and ~70 % of 4 ng 100 bp fragments while recovering ~100 % of 65 ng fragments ranging from 200 bp to 1500 bp.
Figure 7 illustrates an electropherogram (EPG) with 22 out of 24 replicates (outliers excluded) of CleanNGS (left) and AMPure XP (right) when performing double-sided DNA size selection of sheared gDNA using 0.8x-0.7x (left-right) magnetic bead ratios. Both magnetic beads achieved a similar recovery of over 12 % (n=22) of 330 ng sheared gDNA. Average fragment sizes were 372 bp with CleanNGS and 396 bp with AMPure XP magnetic beads, while overall size variation for each reagent was below 5 %.
Remarks
- VIALAB software: The VIALAB programs can be easily adapted to your specific pipette, labware and protocols.
- Partial plates: The pre-set programs offer laboratories complete flexibility to accommodate varying sample sizes, ensuring they can meet both current and future demands.
- Semi-automation: There is a protocol that uses a magnet plate for a semi-automated workflow in the appendix.
Conclusion
- Fully automated DNA size selection and PCR product purification uses the MAG module for reliable magnetic bead handling and the VOYAGER adjustable tip spacing pipette for precise liquid handling on the ASSIST PLUS pipetting robot.
- Double-sided DNA size selection with CleanNGS magnetic beads is game-changing by reducing experimental costs. Fragments of 340 bp to 390 bp are successfully selected with 12 % recovery when using a 0.8x-0.7x (left-right) ratio. On top, CleanNGS results are comparable to AMPure XP, making them the cost-effective alternative.
- Small fragment removal is stress free when performing PCR product purification with a 1.8x CleanNGS magnetic bead ratio, ensuring reproducible recovery of precious fragments above 100 bp.
- Easy adjustment to different protocols with VIALAB's simple programming allows seamless adaptation to changing workloads by modifying magnetic bead volumes and magnet height settings, or sample counts.
Appendix: Complete data set and the semi-automated workflow using a magnet plate
Ask our expert. Leave a comment!
Write us if you have any questions regarding the application note or one of our instruments.
Any questions? I'm happy to help!
Instruments and accessories
ASSIST PLUS, Pipetting Robot
INTEGRA has developed the ASSIST PLUS pipetting robot to streamline routine pipetting tasks at an affordable price. Using INTEGRA electronic multichannel pipettes, the system:
- automates pipetting tasks,
- eliminates physical strain and
- ensures superior reproducibility and
- error free pipetting.
Part No. 4505
VOYAGER, 8 Channel, 125 µl
VOYAGER pipettes allow the tip spacing to expand anywhere between 4.5 mm and 33 mm at the push of a button.
- Single handed operation leaves the other hand free to handle labware.
- On the fly access to up to 3 user-defined tip spacings.
- These user defined tip spacings are saved. No need to memorize spacings of different labware formats.
Part No. 4722
GRIPTIPS®, 125 µl, Sterile, Filter, Low retention
INTEGRA uses a unique polypropylene blend to mold Low Retention GRIPTIPS® with heightened hydrophobic properties. Using Low Retention GRIPTIPS® prevents low surface tension samples from spreading out and “wetting” the inner wall of the tips, allowing them to bead-up for a maximum liquid recovery.
Part No. 6565
MAG module for magnetic separation
- Adjustable magnet heights for bead collection, depending on the volume.
- Exceptional magnet strength, ensuring minimal bead carryover and maximum yields.
- Easy integration with the VIAFLO 96, MINI 96 or VIAFLO 384 pipetting platforms.
- Full automation when combined with the ASSIST PLUS pipetting robot.
Part No. 4900
Irish Life Science: 2.2 ml Square well Pyramid KingFisher plate-type
- Maximum Volume of 2.4ml
- Square well with pyramid bottom for maximum liquid extraction.
- High uniformity from well to well
- Raised well rims for reliable closing with heat sealing.
- Easy and reliable stacking
- Good centrifugation stability up to 6,000 × g for faster protocols and improved sample quality
- Manufactured under DNase/ RNase free environment without slip agents, plasticizers or biocides – Materials which could have a negative effect on bioassays
- Pure, virgin polypropylene guarantees good extractible performance, high resistance to chemicals, good mechanical stress and working with temperature extremes
Source: Website Irish Life Science
BioRad: Hard-Shell 96-well PCR plate, low profile, thin wall, skirted
Instrumentation compatible with low-profile PCR plates include Bio-Rad’s C1000 Touch and S1000 Thermal Cyclers and the CFX series real-time PCR systems. Use the PCR Plastics Consumables Selector Tool to determine individual plate compatibility with other instruments.
Source: Website BioRad
Promega: 100 bp DNA Ladder
The 100bp DNA Ladder has eleven fragments that range in size from 100bp to 1,000bp in 100bp increments with an additional band at 1,500bp. The 500bp fragment is present at increased intensity for easy identification. The ladder is dephosphorylated and ready for 5´ end-labeling with radioisotopes using T4 Polynucleotide Kinase for visualization by autoradiography.
Part Nr.: G2101
Source: Website Promega
CleanNA: CleanNGS
DNA and RNA clean-up for next generation sequencing library construction
The CleanNGS kit offers a highly efficient magnetic bead based clean-up system for the purification of both DNA and/or RNA for next-generation sequencing and sanger sequencing workflows. It can also be used for CRISPR-Cas9. CleanNGS provides maximum flexibility allowing for left, right or double-sided size selection by easily adjusting the sample to CleanNGS volume ratio(s).
Source: Website CleanNA
BECKMAN COULTER Life Sciences: Agencourt AMPure XP
A highly efficient, easily automated PCR purification system that delivers superior quality DNA with no salt carryover. Requiring no centrifugation or filtration, Agencourt AMPure XP can be easily used in manual and automated 96- or 384-well formats.
Source: Website Beckman Coulter
Download
Download App Note as PDF
DownloadAutomated size selection with CleanNGS – the cost-effective alternative to AMPure XPCustomer’s voice
We have to transfer lots of samples and set up sensitive genotyping reactions for national surveillance of malaria and drug resistance. The ASSIST PLUS makes life so EASY and ACCURATE for us! Thank you/Merci.