Specifications

VOYAGER Electronic Pipettes Specifications

VOYAGER, 4 Channel

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

4743

10 – 300

0.5

30
150
300

4.00
2.00
1.60

1.20
0.60
0.35

4744

50 – 1250

1

125
625
1250

6.00
2.40
1.20

1.10
0.50
0.30

VOYAGER, 6 Channel

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

4763

10 – 300

0.5

30
150
300

4.00
2.00
1.60

1.20
0.60
0.35

4764

50 – 1250

1

125
625
1250

6.00
2.40
1.20

1.10
0.50
0.30

VOYAGER, 8 Channel

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

4721

0.5 – 12.5

0.01

1.25
6.25
12.5

10.00
4.00
2.00

6.00
1.60
0.80

4726

2 – 50

0.05

5
25
50

5.00
2.50
1.50

2.50
0.60
0.40

4722

5 – 125

0.1

12.5
62.5
125

3.75
2.50
1.60

1.50
0.70
0.35

4723

10 – 300

0.5

30
150
300

4.00
2.00
1.60

1.20
0.60
0.35

4724

50 – 1250

1

125
625
1250

6.00
2.40
1.20

1.10
0.50
0.30

VOYAGER, 12 Channel

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

4731

0.5 – 12.5

0.01

1.25
6.25
12.5

10.00
4.00
2.00

6.00
1.60
0.80

4736

2 – 50

0.05

5
25
50

5.00
2.50
1.50

2.50
0.60
0.40

4732

5 – 125

0.1

12.5
62.5
125

3.75
2.50
1.60

1.50
0.70
0.35

Remarks

  • The multichannel specifications represent the performance of all channels of a pipette.
  • Precision = Coefficient of variation

 

Download 

Specifications VOYAGER Pipette as PDF