Specifications
VOYAGER Electronic Pipettes Specifications
VOYAGER, 4 Channel
Part No. |
Volume Range |
Volume Increments |
Test Volume |
Accuracy |
Precision |
---|---|---|---|---|---|
4743 |
10 – 300 |
0.5 |
30 |
4.00 |
1.20 |
4744 |
|
1 |
125 |
6.00 |
1.10 |
VOYAGER, 6 Channel
Part No. |
Volume Range |
Volume Increments |
Test Volume |
Accuracy |
Precision |
---|---|---|---|---|---|
4763 |
10 – 300 |
0.5 |
30 |
4.00 |
1.20 |
4764 |
|
1 |
125 |
6.00 |
1.10 |
VOYAGER, 8 Channel
Part No. |
Volume Range |
Volume Increments |
Test Volume |
Accuracy |
Precision |
---|---|---|---|---|---|
4721 |
0.5 – 12.5 |
0.01 |
1.25 |
10.00 |
6.00 |
4726 |
2 – 50 |
0.05 |
5 |
5.00 |
2.50 |
4722 |
5 – 125 |
0.1 |
12.5 |
3.75 |
1.50 |
4723 |
10 – 300 |
0.5 |
30 |
4.00 |
1.20 |
4724 |
|
1 |
125 |
6.00 |
1.10 |
VOYAGER, 12 Channel
Part No. |
Volume Range |
Volume Increments |
Test Volume |
Accuracy |
Precision |
---|---|---|---|---|---|
4731 |
|
0.01 |
1.25 |
10.00 |
6.00 |
4736 |
2 – 50 |
0.05 |
5 |
5.00 |
2.50 |
4732 |
5 – 125 |
0.1 |
12.5 |
3.75 |
1.50 |
Remarks
- The multichannel specifications represent the performance of all channels of a pipette.
- Precision = Coefficient of variation