24 Channel Pipetting Heads

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

6123

10 – 300

0.5

30
150
300

2.50
1.50
1.20

1.50
1.00
0.80

6124

50 – 1250

1

125
625
1250

2.50
2.00
1.80

1.50
1.00
0.80

96 Channel Pipetting Heads

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

6101

0.5 – 12.5

0.01

1.25
6.25
12.5

8.00
2.00
1.50

3.00
1.50
1.00

6106

2 – 50

0.05

5
25
50

5.00
2.00
1.50

2.50
1.50
1.00

6102

5 – 125

0.1

12.5
62.5
125

3.00
2.00
1.50

1.70
1.50
1.00

6103

10 – 300

0.5

30
150
300

2.50
1.50
1.20

1.50
1.00
0.80

6104

50 – 1250

1

125
625
1250

2.50
2.00
1.80

1.50
1.00
0.80

384 Channel Pipetting Heads

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

6131

0.5 – 12.5

0.01

1.25
6.25
12.5

8.00
2.00
1.50

3.00
1.50
1.00

6136

2 – 50

0.05

5
25
50

5.00
2.00
1.50

2.50
1.50
1.00

6132

5 – 125

0.1

12.5
62.5
125

3.00
2.00
1.50

1.70
1.50
1.00

Specifications represent the performance of all channels, specifications apply to neat transfers in Pipet mode only, precision = coefficient of variation