Automated DNA extraction protocol to effortlessly isolate gDNA from blood

  • Automated DNA extraction protocol to effortlessly isolate gDNA from blood

    Hands-free high yielding gDNA purification with magnetic beads

    Genomic DNA (gDNA) purification with magnetic beads provides a robust and versatile method for isolating nucleic acids from various biological samples in molecular biology workflows. Extracting DNA from blood enables advanced genomic studies across diverse fields such as oncology, hematology or immunology. The Clean Blood & Tissue DNA kit is specifically designed to isolate gDNA for highly efficient nucleic acid purification, and is compatible with fully automated DNA extraction protocols.

    This application note demonstrates INTEGRA's automated DNA extraction protocol, providing a walk-away solution for extracting gDNA from 12 EDTA blood samples. It relies on the VOYAGER adjustable tip spacing pipette to ensure reliable liquid handling of multiple tubes in parallel on the ASSIST PLUS pipetting robot. In addition, the HEATMAG module facilitates automated magnetic bead handling and guarantees temperature control for improved lysis and elution performance.

    The protocol provides reliable, high yielding gDNA purification with magnetic beads using the Clean Blood & Tissue DNA Kit, optimizing gDNA quality for downstream applications.

  • Table of contents

    Genomic DNA (gDNA) purification with magnetic beads provides a robust and versatile method for isolating nucleic acids from various biological samples in molecular biology workflows. Extracting DNA from blood enables advanced genomic studies across diverse fields such as oncology, hematology or immunology. The Clean Blood & Tissue DNA kit is specifically designed to isolate gDNA for highly efficient nucleic acid purification, and is compatible with fully automated DNA extraction protocols.

    This application note demonstrates INTEGRA's automated DNA extraction protocol, providing a walk-away solution for extracting gDNA from 12 EDTA blood samples. It relies on the VOYAGER adjustable tip spacing pipette to ensure reliable liquid handling of multiple tubes in parallel on the ASSIST PLUS pipetting robot. In addition, the HEATMAG module facilitates automated magnetic bead handling and guarantees temperature control for improved lysis and elution performance.

    The protocol provides reliable, high yielding gDNA purification with magnetic beads using the Clean Blood & Tissue DNA Kit, optimizing gDNA quality for downstream applications.

Key benefits

  • The automated DNA extraction protocol for the Clean Blood & Tissue DNA Kit enables the extraction of pure, high yielding gDNA from 250 µl of EDTA blood.
  • Automated gDNA purification with magnetic beads – combined with the unique features of the HEATMAG module – guarantees optimal results, with automated magnet engagement during washing, and temperature control during lysis and elution.
  • Using the ASSIST PLUS for liquid handling steps frees up the user for other tasks when processing small sample batches.
  • Flexible throughput of up to 24 samples is supported by optimized VIALAB programs, ensuring consistent extraction performance across different batch sizes.
  • Quick start guides help operators to quickly set up the application for 6, 12, 18 or 24 samples, and can be downloaded below together with the corresponding VIALAB programs.
  • The ASSIST PLUS is designed to fit in a biosafety cabinet, allowing users to safely perform the high yielding gDNA purification workflow with hazardous samples.

Overview: How to automate gDNA extraction from blood on the ASSIST PLUS

This application note presents a comprehensive automated DNA extraction protocol for 12 EDTA blood samples in 1.5 ml microcentrifuge tubes using the Clean Blood & Tissue DNA Kit on the ASSIST PLUS. The protocol guarantees precise automated liquid handling with the 50-1,250 µl 6 channel VOYAGER, while the HEATMAG module offers automated magnetic bead handling with temperature control.

The step-by-step procedure for the automated DNA extraction protocol is illustrated in Figure 1.

Illustration of the step-by-step procedure of blood extraction and gDNA purification with magnetic beads.
Figure 1: Step-by-step illustration of blood extraction and gDNA purification with magnetic beads.

Downloads: App note, Quick start guides and protocols for automated DNA extraction protocol to effortlessly isolate gDNA from blood

Experimental set-up

Graphical representation of the ASSIST PLUS deck set-up with an INTEGRA DWP, HEATMAG module with tube adapter, Slider base plate with Sliders and the VOYAGER.
Figure 2: ASSIST PLUS deck set-up for the automated DNA extraction protocol for 12 EDTA blood samples. Position A: INTEGRA deep well plate (DWP) (light blue: 670 µl lysis mix; green: 550 µl Blood DNA Particle Solution; lilac: 1,300 µl GH Wash; red: 1,300 µl TB Wash; orange: 340 µl Elution Buffer. Position B: HEATMAG module with 12 sample tubes (blue) in positions A1-B6 Position C: Sliders C1 and C2 with 12x1.5 ml microcentrifuge tubes for elution (pink).

Step-by-step procedure

1. Lysis

Lyse blood cells

Prepare the lysis mix as indicated in the user manual (290 µl TN Lysis + 20 µl Proteinase K). Multiply the single components for the lysis mix by the sample count, adding 2 extra reactions as dead volume (e.g. 14 for a sample count of 12). Add the prepared lysis mix and the other reagents from the Clean Blood & Tissue DNA Kit to a new INTEGRA DWP. Add 250 µl of thawed EDTA Blood to 12 new 1.5 ml microcentrifuge tubes and set up the ASSIST PLUS deck according to Figure 2.

Select and run the VIALAB program 'DNA_extraction_from_blood_12samples'. The HEATMAG module will start heating to reach 56 °C. Using 1,250 µl sterile, filter GRIPTIPS® pipette tips, the VOYAGER on the ASSIST PLUS will transfer 310 µl of the lysis mix to each EDTA blood sample. During lysis, samples are mixed at Speed 3 to avoid spillage when handling hazardous materials. Optimal lysis conditions are enabled by alternated mixing of column A and B, which ensures thorough homogenization of each sample.

Tip:

  • Preheat the HEATMAG module to 56 °C in advance to reduce the total run time.

2. Binding

Bind gDNA to magnetic beads

The HEATMAG module will start active cooling until ambient temperature is reached. Simultaneously, the VOYAGER will mix the Blood DNA Particle Solution 5 times in the DWP to ensure homogeneity before transferring 250 µl to each lysate on the HEATMAG module (Figure 3). After adding the Blood DNA Particle Solution, the samples are mixed continuously to ensure efficient binding of gDNA to the magnetic beads.

The ASSIST PLUS and VOYAGER transfer Blood DNA Particle Solution containing magnetic beads from the INTEGRA DWP into tubes containing lysed whole blood on the HEATMAG module.
Figure 3: The VOYAGER on the ASSIST PLUS transfers the Blood DNA Particle Solution from (a) the DWP with reagents to (b) the 1.5 ml microcentrifuge tubes containing lysates on the HEATMAG module equipped with a 24 position tube adapter.

3. Washing

Wash gDNA bound to magnetic beads

The HEATMAG module will automatically move the magnet array from the home position to the high position (Figure 4) for 10 minutes to capture the magnetic beads (Figure 5). The VOYAGER then removes the supernatant while the magnet array remains engaged, before returning to the home position.

To initiate the first wash cycle, the VOYAGER transfers 600 µl of GH Wash to each microcentrifuge tube on the HEATMAG module. To achieve optimal washing conditions, columns A and B are alternately mixed in two stages. To detach the magnetic beads from the tube wall, the first round of mixing is performed with static pipette tips close to the bead pellet (x=y= 1.8 mm offset). For the second mix, the magnetic beads are resuspended using tip-travel with x=y= -1.2 mm offset.

The operator is prompted to change tip box and waste bin. The new tip box is initialized once the operator confirms the tip dialogue. The HEATMAG module then captures the magnetic beads for 1 minute in the high position. The ASSIST PLUS repeats the wash cycle using TB Wash. A slow aspiration (Speed 1) and precise height settings away from the pellet (x=y= -1.2 mm offset) prevent bead loss when removing the supernatant while washing.

At the end of the second wash cycle, the HEATMAG module will lift the magnet array to the low position at 24 mm to capture the magnetic beads closer to the bottom of the microcentrifuge tube. After 1 minute, the VOYAGER removes the supernatant before the HEATMAG module lowers the magnet array to the elution height of 21 mm. Following another 1-minute incubation, the VOYAGER will aspirate any remaining TB Wash and initialize a 10-minute drying step with the HEATMAG module.

Tip:

  • Set the timer to 40 minutes at the start of the run to ensure the tip box and waste bin are changed on time.
The HEATMAG module on the ASSIST PLUS without a tube adapter, showing both the disengaged magnet array in the home position, and the engaged magnet array in the high position.
Figure 4: The HEATMAG module on the ASSIST PLUS without a tube adapter, showing (a) the disengaged magnet array (home position, 0 mm) and (b) the engaged magnet array (high position, 29 mm).
The HEATMAG module on the ASSIST PLUS with a tube adapter and 1.5 ml microcentrifuge tubes showing both uncaptured and captured magnetic beads.
Figure 5: The HEATMAG module on the ASSIST PLUS with a tube adapter and 1.5 ml microcentrifuge tubes, showing (a) uncaptured and (b) captured magnetic beads in positions A1-B6.

4. Elution

Elute gDNA from magnetic beads

The HEATMAG module will lower the magnet array back to the home position, then the VOYAGER will transfer 120 µl of Elution Buffer to each microcentrifuge tube on the HEATMAG module. A pre-dispense step ensures precise transfers when pipetting below 10 % of the nominal volume. Slow dispensing of Elution Buffer directly on the bead pellet (x=y= +1.8 mm offset) ensures proper rehydration of the magnetic beads. Every sample is mixed 40 times by aspirating and dispensing 100 µl and incubated for 10 minutes at 65 °C. Following incubation, the HEATMAG module lifts the magnet to the low position at 24 mm to capture the magnetic beads for 10 minutes. The VOYAGER will then aspirate 110 µl of the eluates at Speed 1 to prevent magnetic bead carryover, transferring them to the corresponding empty 1.5 ml microcentrifuge tubes in Sliders C1 and C2 (Figure 2; pink). The operator will be informed when the run is complete.

Tip:

  • The elution volume can be increased to 200 µl if sample dilution is preferred.

Results

Insufficient gDNA purification from whole blood can impact downstream molecular biology applications such as qPCR and NGS. This application note demonstrates the effectiveness of the Clean Blood & Tissue DNA Kit in combination with INTEGRA's automated DNA extraction protocol, optimized for the HEATMAG module on the ASSIST PLUS, using the 50-1,250 µl 6 channel VOYAGER.

A blood bank provided EDTA blood samples, which were stored at 4 °C for no more than 3 days. Tubes from 10 different donors with the same blood group were pooled, aliquoted, and frozen until use. Each run included 12 microcentrifuge tubes containing 240 µl of blood and 10 µl of DNA spike-in control to verify qPCR performance and exclude possible downstream qPCR inhibition. DNA yield and purity were analyzed via spectrophotometry (DeNovix  DS-11 Series), and qPCR performance was assessed using the QuantStudio 3 (Thermo Fisher Scientific).

The performance of the Clean Blood & Tissue DNA Kit on the ASSIST PLUS was assessed by measuring the yields, purity ratios and qPCR analysis of 1 in 10 dilutions of the eluate (Figure 6). On average, the automated DNA extraction protocol yielded 5.4 µg of gDNA, with an average purity ratio of 1.9 for A260/280 and 2.0 for A260/230 from the specific blood batch (Figure 6 a-c). Furthermore, qPCR analysis shows no signs of inhibition, as inferred from the consistent cycle threshold (Ct) increase of 3.3 and 3.5 for the 1x–10x and 10x–100x diluted input DNA. The consistent qPCR performance across samples confirms the robustness and reliability of the automated DNA extraction protocol (Figure 6 d-e).

High yielding gDNA purification from blood with the Clean Blood & Tissue DNA Kit. Error bars indicate the standard deviation (n=36). Detailed data can be found in the appendix. (a) Bar chart of total DNA yield. (b) Bar chart of OD values for A260/280 ratios. (c) Bar chart of OD values for A260/230 ratios. (d) Bar chart of Ct values. (e) Amplification plot of a single 12 sample run showing undiluted (green), 10-fold diluted (light blue) and 100-fold diluted (orange) eluates.
Figure 6: High yielding gDNA purification from blood with the Clean Blood & Tissue DNA Kit. Error bars indicate the standard deviation (n=36). Detailed data can be found in the appendix. (a) Total DNA yield. (b) OD values for A260/280 ratios. (c) OD values for A260/230 ratios. (d) Ct values. (e) Amplification plot of a single 12 sample run showing undiluted (green), 10-fold diluted (light blue) and 100-fold diluted (orange) eluates.

Remarks

  • VIALAB software: VIALAB programs can be quickly adapted to the user’s specific pipette, labware, modules and protocols.
  • HEATMAG module: The magnet step in VIALAB enables precise control over the magnet array, allowing users to set customized heights between 0 and 29 mm. In addition, the temperature can be controlled through VIALAB for each step, ranging from room temperature to 65 °C.

Conclusion

  • The high yielding gDNA purification workflow enabled the extraction of ~5.4 µg gDNA from 240 µl EDTA blood using the automated DNA extraction protocol for the Clean Blood & Tissue DNA Kit on the ASSIST PLUS with the HEATMAG module.
  • Precise liquid and magnetic bead handling with the VOYAGER and HEATMAG module on the ASSIST PLUS ensure reliable results, guaranteeing high data quality.
  • The purity of the extracted gDNA – indicated by an A260/280 ratio of ~1.9 and A260/230 ratio of ~2 – ensures optimal qPCR performance, due to the absence of contaminants such as proteins or organic contaminants (e.g. phenol, guanidine, salts).
  • Automated processing of up to 24 samples reduces hands-on time and ensures an efficient workflow, allowing small batches to be processed immediately upon arrival of samples, reducing time-to-results.
  • Highly consistent extraction performance across different sample throughputs is supported by optimized VIALAB programs. Detailed data for the processing of 6, 12, 18 and 24 samples is provided in the appendix.

Ask our expert. Leave a comment!

Write us if you have any questions regarding the application note or one of our instruments.

Any questions? I'm happy to help!

  • Yvo Furrer

    Lab Specialist

    Email

Instruments and accessories

ASSIST PLUS, Pipetting Robot Slider base plate for 4 sliders Slider for 1.5/2 ml microcentrifuge tubes (D-ONE compatible) VOYAGER, 6 Channel, 1250 µl, electronic pipette HEATMAG module Adapter for 1.5 ml tubes for MAG and HEATMAG GRIPTIPS®, 1250 µl, Sterile, Filter INTEGRA 96 square well, 2.2 ml deep well plate Clean Blood & Tissue DNA Kit 1.5 ml microcentrifuge tubes (Greiner Bio-One) TaqMan™ Universal DNA Spike-In Control EDTA blood samples

ASSIST PLUS, Pipetting Robot

INTEGRA has developed the ASSIST PLUS pipetting robot to streamline routine pipetting tasks at an affordable price. Using INTEGRA electronic multichannel pipettes, the system:

  • automates pipetting tasks,
  • eliminates physical strain and
  • ensures superior reproducibility and
  • error free pipetting.

Part No. 4505

Visit product page
ASSIST PLUS Pipetting Robot

Slider base plate for 4 sliders

  • Slider base plate to accomodate up to 4 different sliders
  • Sliders can be exchanged for alsmost unlimited labware flexibility
  • Individual sliders allows a contamination free loading of labware

Part No. 4561

slider base plate for 4 sliders

Slider for 1.5/2 ml microcentrifuge tubes (D-ONE compatible)

  • Compatible with 4, 6 or 8 channel VOYAGER adjustable tip spacing pipettes or D-ONE
  • Stores up to 16 x 1.5 or 2 ml microcentrifuge tubes (screw cap tubes or with attatched lid)

Part number: 4562

slider for 1.5 2ml microcentrifuge tubes, compatible with d-one

VOYAGER, 6 Channel, 1250 µl, electronic pipette

VOYAGER pipettes allow the tip spacing to expand anywhere between 4.5 mm and 33 mm at the push of a button.

  • Single handed operation leaves the other hand free to handle labware.
  • On the fly access to up to 3 user-defined tip spacings.
  • These user defined tip spacings are saved. No need to memorize spacings of different labware formats.

Part No. 4764

See it work
VOYAGER, 6 channels, 1250μL

HEATMAG module

  • Vertical magnet movement for automated magnetic bead purifications
  • Strong magnets for fast bead collection
  • Allows for heating up from ambient to 65°C for lysis or elution
  • Can be used stand-alone or on the ASSIST PLUS, VIAFLO 96, VIAFLO 384 or MINI 96
  • Includes module, 24 disc magnetic array, power supply, communication cable to ASSIST PLUS, communication cable to PC
  • No labware adapter included
  • For communication with MAG Control app install the communication module PN 4222

Part number: 4901

 

Visit product page
heatmag module without adapter

Adapter for 1.5 ml tubes for MAG and HEATMAG

  • Compatible with MAG and HEATMAG modules
  • For 1.5 ml microcentrifuge tubes with snap caps or screw caps
  • Contains cut-outs to hold snap caps
  • Accommodates 24 tubes in format 4 x 6

Part number: 4905

Visit product page
heatmag adapter for 1.5 ml tubes

GRIPTIPS®, 1250 µl, Sterile, Filter

GRIPTIPS® pipette tips perfectly match the multi-lobe tip fitting, snapping firmly on during loading. This guarantees a perfect seal on every tip, preventing them from loosening, leaking or completely falling off. All tips are precisely aligned horizontally, enabling accurate touch-offs, even when pipetting with 384 tips.

Part No. 6445

Learn more about GRIPTIPS® pipette tips for Benchtop Pipetting Systems
Integra-Relaunch_Website-Output-Produktebilder-Product_Overview-GripTips.png

INTEGRA 96 square well, 2.2 ml deep well plate

  • Maximum Volume of 2.4ml
  • Square well with pyramid bottom
  • Raised well rims for reliable closing with heat sealing and easy stacking
  • Good centrifugation stability up to 6,000 × g for faster protocols and improved sample quality
  • DNase/ RNase free, without slip agents, plasticizers or biocides

Part No. 6353

Learn more about our deep well plate
INTEGRA 96 square well, V bottom, polypropylene, sterile, 2.2 ml deep well plates

Clean Blood & Tissue DNA Kit

  • Extraction kit for whole blood, buffy coat, cultured cells and up to 10 mg of tissue
  • Fully compatible with blood samples in EDTA, citrate and heparin tubes
  • Extracted DNA is immediately ready for downstream applications

Part No. 7100 / 7101 / 7102

Visit product page
Clean Blood & Tissue Kit

1.5 ml microcentrifuge tubes (Greiner Bio-One)

  • Free of detectable DNase, RNase, human DNA
  • Non-Pyrogenic
  • High chemical and temperature resistance
  • Volume: 1.5 ml
  • With graduation and attached cap
  • Suitable for Eppendorf

Part No. 616201

Source: Website Supplier

Visit supplier website
Reaction tube, 1,5 ml, PP, natural

TaqMan™ Universal DNA Spike-In Control

TaqMan Universal DNA Spike-In Control serves as an exogenous process control of a known concentration that is spiked into a sample at specific stages throughout the qPCR process. Measuring the concentration of the spiked-in sequence allows for its use as a serial quality control and it can be used to monitor recovery efficiency and to identify false negative results such as the presence of PCR inhibitors in molecular detection workflows. This is of particular importance when working with sample types that have a high level of inhibition.

Note: this product contains a sequence for AmpC beta-lactamase and may amplify for any assays designed to detect this antibiotic resistance target.

Part No. A39175

Source: Website Supplier

Visit supplier website

EDTA blood samples

  • Whole blood samples from blood donors with one blood type
  • Blood is tested on common blood-borne infectious diseases
  • The blood samples contain EDTA as an anticoagulant

Part No. 92000

Source: Website Supplier

Visit supplier website

Customer’s voice

Since we have implemented the Assist Plus into our workflow, we have been able to streamline our lab processes. It has enabled us to do more in the lab and with accuracy. The instrument can be standalone or with a PC. The software is easy to understand and use. I highly recommend this product to increase throughput.