Specifications EVOLVE Pipettes

EVOLVE Manual Pipettes Specifications

EVOLVE, Single Channel

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

3011

0.2 – 2.0

0.001

0.2
1
2

12.0
3.0
1.5

6.0
1.6
0.7

3012

1 – 10

0.01

1
5
10

5.0
1.5
1.2

1.5
0.8
0.5

3013

2 – 20

0.01

2
10
20

7.0
1.6
1.0

2.0
0.5
0.35

3015

10 – 100

0.1

10
50
100

3.5
1.0
0.8

1.0
0.35
0.2

3016

20 – 200

0.1

20
100
200

2.5
1.0
0.7

1.0
0.3
0.18

3018

100 – 1000

1

100
500
1000

3.0
1.0
0.8

0.6
0.3
0.18

3019

500 – 5000

5

500
2500
5000

3.0
1.2
0.6

0.6
0.3
0.18

EVOLVE, 8 Channel

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

3022

1 – 10

0.01

1
5
10

12.0
4.0
2.0

7.0
3.0
1.2

3024

5 – 50

0.05

5
25
50

5.0
2.0
1.2

2.0
1.0
0.5

3025

10 – 100

0.1

10
50
100

5.0
2.0
1.2

1.5
0.8
0.3

3026

20 – 200

0.1

20
100
200

5.0
1.8
1.0

1.2
0.5
0.3

3028

100 – 1250

1.0

125
625
1250

5.0
1.8
1.0

1.1
0.7
0.3

EVOLVE, 12 Channel

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

3032

1 – 10

0.01

1
5
10

12.0
4.0
2.0

7.0
3.0
1.2

3034

5 – 50

0.05

5
25
50

5.0
2.0
1.2

2.0
1.0
0.5

3035

10 – 100

0.1

10
50
100

5.0
2.0
1.2

1.5
0.8
0.3

3036

20 – 200

0.1

20
100
200

5.0
2.0
1.2

1.2
0.5
0.3

3038

100 – 1250

1

125
625
1250

5.0
1.8
1.0

1.1
0.7
0.3

EVOLVE, 16 Channel

Part No. 

Volume Range
(µl)

Volume Increments 
(µl)

Test Volume
(µl)

Accuracy
(±%)

Precision
(≤%)

3042

1 – 10

0.01

1
5
10

12.0
4.0
2.0

7.0
3.0
1.2

3044

5 – 50

0.05

5
25
50

5.0
2.0
1.2

2.0
1.0
0.5

3045

10 – 100

0.1

10
50
100

5.0
2.0
1.2

1.5
0.8
0.3

Remarks

  • The multichannel specifications represent the performance of all channels of a pipette.
  • Precision = Coefficient of variation

 

Download 

Specifications EVOLVE Pipette as PDF