Specifications EVOLVE Pipettes
EVOLVE Manual Pipettes Specifications
EVOLVE, Single Channel
Part No. |
Volume Range |
Volume Increments |
Test Volume |
Accuracy |
Precision |
---|---|---|---|---|---|
3011 |
0.2 – 2.0 |
0.001 |
0.2 |
12.0 |
6.0 |
3012 |
1 – 10 |
0.01 |
1 |
5.0 |
1.5 |
3013 |
2 – 20 |
0.01 |
2 |
7.0 |
2.0 |
3015 |
10 – 100 |
0.1 |
10 |
3.5 |
1.0 |
3016 |
20 – 200 |
0.1 |
20 |
2.5 |
1.0 |
3018 |
|
1 |
100 |
3.0 |
0.6 |
3019 |
500 – 5000 |
5 |
500 |
3.0 |
0.6 |
EVOLVE, 8 Channel
Part No. |
Volume Range |
Volume Increments |
Test Volume |
Accuracy |
Precision |
---|---|---|---|---|---|
3022 |
1 – 10 |
0.01 |
1 |
12.0 |
7.0 |
3024 |
5 – 50 |
0.05 |
5 |
5.0 |
2.0 |
3025 |
10 – 100 |
0.1 |
10 |
5.0 |
1.5 |
3026 |
20 – 200 |
0.1 |
20 |
5.0 |
1.2 |
3028 |
|
1.0 |
125 |
5.0 |
1.1 |
EVOLVE, 12 Channel
Part No. |
Volume Range |
Volume Increments |
Test Volume |
Accuracy |
Precision |
---|---|---|---|---|---|
3032 |
1 – 10 |
0.01 |
1 |
12.0 |
7.0 |
3034 |
5 – 50 |
0.05 |
5 |
5.0 |
2.0 |
3035 |
10 – 100 |
0.1 |
10 |
5.0 |
1.5 |
3036 |
20 – 200 |
0.1 |
20 |
5.0 |
1.2 |
3038 |
|
1 |
125 |
5.0 |
1.1 |
EVOLVE, 16 Channel
Part No. |
Volume Range |
Volume Increments |
Test Volume |
Accuracy |
Precision |
---|---|---|---|---|---|
3042 |
|
0.01 |
1 |
12.0 |
7.0 |
3044 |
5 – 50 |
0.05 |
5 |
5.0 |
2.0 |
3045 |
10 – 100 |
0.1 |
10 |
5.0 |
1.5 |
Remarks
- The multichannel specifications represent the performance of all channels of a pipette.
- Precision = Coefficient of variation
Download
Specifications EVOLVE Pipette as PDF