Specifications EVOLVE Pipettes
EVOLVE Manual Pipettes Specifications
EVOLVE, Single Channel
Part No. | Volume Range | Volume Increments | Test Volume | Accuracy | Precision |
---|---|---|---|---|---|
3011 | 0.2 – 2.0 | 0.001 | 0.2 | 12.0 | 6.0 |
3012 | 1 – 10 | 0.01 | 1 | 5.0 | 1.5 |
3013 | 2 – 20 | 0.01 | 2 | 7.0 | 2.0 |
3015 | 10 – 100 | 0.1 | 10 | 3.5 | 1.0 |
3016 | 20 – 200 | 0.1 | 20 | 2.5 | 1.0 |
3018 | 1 | 100 | 3.0 | 0.6 | |
3019 | 500 – 5000 | 5 | 500 | 3.0 | 0.6 |
EVOLVE, 8 Channel
Part No. | Volume Range | Volume Increments | Test Volume | Accuracy | Precision |
---|---|---|---|---|---|
3022 | 1 – 10 | 0.01 | 1 | 12.0 | 7.0 |
3024 | 5 – 50 | 0.05 | 5 | 5.0 | 2.0 |
3025 | 10 – 100 | 0.1 | 10 | 5.0 | 1.5 |
3026 | 20 – 200 | 0.1 | 20 | 5.0 | 1.2 |
3028 | 1.0 | 125 | 5.0 | 1.1 |
EVOLVE, 12 Channel
Part No. |
Volume Range |
Volume Increments |
Test Volume |
Accuracy |
Precision |
---|---|---|---|---|---|
3032 |
1 – 10 |
0.01 |
1 |
12.0 |
7.0 |
3034 |
5 – 50 |
0.05 |
5 |
5.0 |
2.0 |
3035 |
10 – 100 |
0.1 |
10 |
5.0 |
1.5 |
3036 |
20 – 200 |
0.1 |
20 |
5.0 |
1.2 |
3038 |
|
1 |
125 |
5.0 |
1.1 |
EVOLVE, 16 Channel
Part No. |
Volume Range |
Volume Increments |
Test Volume |
Accuracy |
Precision |
---|---|---|---|---|---|
3042 |
|
0.01 |
1 |
12.0 |
7.0 |
3044 |
5 – 50 |
0.05 |
5 |
5.0 |
2.0 |
3045 |
10 – 100 |
0.1 |
10 |
5.0 |
1.5 |
Remarks
- The multichannel specifications represent the performance of all channels of a pipette.
- Precision = Coefficient of variation
Download
Specifications EVOLVE Pipette as PDF