Automated RT-PCR set-up for COVID-19 testing
How to prepare RT-PCR plates for SARS-CoV-2 detection with the ASSIST PLUS
The emergence and outbreak of the novel coronavirus SARS-CoV-2 (COVID-19) has placed unprecedented demands on laboratories testing for COVID-19, leaving scientific staff to contend with a spiraling influx of patient samples and a rapid, continuous growth in workload. Laboratories need additional automated liquid handling instruments for viral nucleic acid extraction and RT-PCR set-up – which are the most labor-intensive processes in the diagnostic testing workflow – to increase the sample throughput capacity, reduce manual intervention by laboratory analysts and fast track the development of COVID-19 assays.
The ASSIST PLUS pipetting robot together with a VOYAGER adjustable tip spacing pipette, Low Retention GRIPTIPS® and SureFlo™ 10 ml Reagent Reservoirs were successfully used for RT-PCR set-up in COVID-19 testing laboratories.
-
Table of contents
The emergence and outbreak of the novel coronavirus SARS-CoV-2 (COVID-19) has placed unprecedented demands on laboratories testing for COVID-19, leaving scientific staff to contend with a spiraling influx of patient samples and a rapid, continuous growth in workload. Laboratories need additional automated liquid handling instruments for viral nucleic acid extraction and RT-PCR set-up – which are the most labor-intensive processes in the diagnostic testing workflow – to increase the sample throughput capacity, reduce manual intervention by laboratory analysts and fast track the development of COVID-19 assays.
The ASSIST PLUS pipetting robot together with a VOYAGER adjustable tip spacing pipette, Low Retention GRIPTIPS® and SureFlo™ 10 ml Reagent Reservoirs were successfully used for RT-PCR set-up in COVID-19 testing laboratories.
Key benefits
Overview: Automated RT-PCR set-up
The ASSIST PLUS pipetting robot is used to automate testing of suspected COVID-19 positive cases in a 384 well plate. The pipetting robot operates a VOYAGER 12 channel 50 μl electronic pipette with 125 μl Sterile, Filter, Low Retention GRIPTIPS. To double the available testing capacity and, concurrently, decrease the cost per test of expensive One-Step RT-PCR reagents of dwindling availability, the total PCR reaction volume was miniaturized, reducing it to 10 μl – inclusive of 7.5 μl Onestep RT-PCR Mastermix and 2.5 μl of nucleic acid template. The templates were extracted from combined nasopharyngeal/ oropharyngeal flocked swabs or sputum samples. The following procedure is based on the protocol used by the Microbiology and Molecular Pathology Department at Sullivan Nicolaides Pathology (SNP) – part of the Sonic Healthcare Group – in Brisbane, Australia.
The protocol is divided into two parts:
- Program 1: Add the Mastermix (1-COVID-19)
- Program 2: Add the nucleic acid template (2-COVID-19)
Downloads: Protocols and reports for automated RT-PCR testing for COVID-19 with ASSIST PLUS
Here you can download the complete protocols and reports used in this application and use them on your ASSIST PLUS. The protocol can be customized with the VIALAB software.
Experimental set-up: Program 1
|
Deck position A: |
10 ml reagent reservoir with SureFlo anti-sealing array containing 3 ml of One-step RT-PCR Mastermix. |
| Deck position C: |
384 well plate placed on a PCR 384 well cooling block, allowing Mastermix and samples to be kept cold, and enabling exact positioning of the PCR plate on the deck. |
Step-by-step procedure
1. Add the Mastermix
Fill the 384 well plate with the One-step RT-PCR Mastermix.
Place the One-step RT-PCR Mastermix in a 10 ml sterile, polystyrene reagent reservoir with INTEGRA’s SureFlo anti-sealing array. Set up the deck with the required labware, as indicated in Figure 1. Select the VIALAB program 1-COVID-19. The VOYAGER pipette automatically transfers the Mastermix from the reservoir into the 384 well plate (LightCycler® 480 Multiwell Plate, Roche) using the Repeat Dispense Mode with tip touch. Each well of the plate is filled with 7.5 μl of Mastermix.
Tips:
- Using a 10 ml reagent reservoir with SureFlo anti-sealing array allows a very low dead volume (<20 μl) to minimize the loss of expensive reagent of dwindling availability (see Figure 2).
- The combination of a low pipetting speed – set at 2 – and Low Retention GRIPTIPS shows excellent results when pipetting the viscous and foamy Mastermix.
- Pre- and post-dispense settings, together with the tip touch option, guarantee reproducible, precise and accurate pipetting results (see Figure 2).
- The PCR cooling block is used as a support to fix the position of the 384 well plate on the deck, ensuring exact tip positioning when pipetting. The cooling block also helps to keep samples and reagents cool if required by the protocol.
Experimental set-up: Program 2
|
Deck position A and B: |
FluidX Cluster 0.7 ml tubes containing the nucleic acid templates. The tubes are stored in a 96-format rack. A total of four sample racks are used for the protocol (two on Position A and two on Position B). |
| Deck position C: |
384 well plate placed on a PCR 384 well cooling block. |
Step-by-step procedure
2. Add the nucleic acid templates
Transfer the samples from four 96-format tube racks to the 384 well plate.
Nucleic acid templates extracted from combined nasopharyngeal/oropharyngeal flocked swabs or sputum samples are stored in FluidX Cluster 0.7 ml tubes placed in a 96-format rack. The VOYAGER pipette transfers 2.5 μl of template from the tubes to the 384 well plate, automatically changing the GRIPTIP pipette tips after each dispense. Both Position A and B are used to house the samples on the deck (see Figure 3). The pipette prompts the user when it is time to replace the tube racks on the deck. After user confirmation, the VOYAGER pipette continues reformatting the samples from tubes to the plate.
Tips:
- The VOYAGER pipette’s tip spacing capability combined with automatic Tip Change ensures easy and rapid sample transfer without risk of contamination or reformatting errors.
- Using an air gap of 1.5 μl when aspirating the viral nucleic acid template eliminates the risk of contamination risk during pipette tip travel.
Note: Automated RT-PCR testing for COVID-19 with the ASSIST PLUS can also be performed using a VOYAGER 8 channel 50 μl electronic pipette (see Figure 5).
Remarks
4 Position Portrait Deck:
If your process allows, the protocol can be compiled in one simple program using the 4 Position Portrait Deck option of the ASSIST PLUS (see Figure 6).
96 well plates:
The protocol can be readily adapted to 96 well format.
VIALAB software:
The VIALAB programs can be easily adapted to your specific labware and protocols.
For more information about the use of the ASSIST PLUS for COVID-19 sample testing and RT-PCR set-up at the Microbiology and Molecular Pathology Department at Sullivan Nicolaides Pathology, please click here.
Conclusion
Related articles
Ask our expert. Leave a comment!
Write us if you have any questions regarding the application note or one of our instruments.
Any questions? I'm happy to help!
Instruments and accessories
ASSIST PLUS, Pipetting Robot
INTEGRA has developed the ASSIST PLUS pipetting robot to streamline routine pipetting tasks at an affordable price. Using INTEGRA electronic multichannel pipettes, the system:
- automates pipetting tasks,
- eliminates physical strain and
- ensures superior reproducibility and
- error free pipetting.
Part No. 4505
Customer’s voice
Easy and flexible pipetting robot! The device is easy to use and therfore it saves us a lot of work. In addition, pipetting is more precise that manual pipetting and there is less risk of contamination. The software for making protocols is very inuitive and provides a lot of options, so the Assist can be easily used for different applications.
Related products
Applications
Show all-
End-to-end Twist library preparation protocol for NGS workflow automation
-
A simple and streamlined workflow for membrane protein isolation
-
Streamlining single-cell sequencing with the Flex Apex workflow using automated liquid handling solutions
-
Optimizing DNA fragmentation and NGS library preparation with automated liquid handling
Stories
Show all-
· Customer TestimonialEnhancing the consistency of ag-tech R&D processes with the ASSIST PLUS pipetting robot
-
· Product NewsMagnetic beads represent a modern approach to gDNA isolation
-
· Product NewsMagnetic beads for NGS size selection and PCR purification
-
· Customer TestimonialGenerating the largest scRNA-seq datasets in the world with automated pipetting